The tables list each study type's columns: the use a column is given under Adjust columns in Import a file, the matching column in the Data grid, whether the study needs it, and what to enter. See also Importing a file, Numbers, units and identifiers and Templates and downloads.

Fields every study type offers#

These columns are optional.

Column useWhat to enter
Excludedyes/no, true/false or 1/0, any case. Leave empty to include the row. An excluded row needs a reason
Exclusion reasonFree text, up to 500 characters
UnitMust match the study's unit exactly. Rows in other units form separate groups
AnalyteCompared with the study's analyte
Sample typeFor example Serum. Compared with the study's matrix
InstrumentCompared with the study's instrument

Unit, Analyte, Matrix and Instrument sort rows into groups. This study gets the rows that match it, and other groups can become new studies. See Files that mix analytes, matrices, instruments or units. Qualitative agreement has no unit, so delete any Unit column before importing. In instrument comparison, Instrument is required.

When the file has no level, day or run column, each is set to 1. Other missing columns, the level in linearity, and the day and run in a bias check stay blank.

Headings the app recognizes#

A matching heading is read as that column use. Case and a trailing unit in brackets are ignored, so Result (mg/dL) counts as result. A heading that contains a recognized word, in English, German or French (for example Ergebnis or Probe), is read the same way. Numbered columns such as Rep1 and Rep2 are offered as one result column with a row per value. A heading that can mean several things (x, y, test, reference, id, sample, value, concentration, time, factor, group, sex, system or reason) is offered as a suggestion for you to confirm. If two columns match one use, you choose.

FieldRecognized headings
IDspecimen, specimen id, specimen_id, material id, material_id, id, sample, sample id, subject, subject id, pool, pool id, specimen / material id
Resultresult, value, measurement, measured, measured result, measured value, result category
Comparative method resultcomparator, comparator result, comparator_result, reference, reference result, x, comparative, comparative result, comparative method result
Test method resultcandidate, candidate result, candidate_result, test, test result, y, test method result
Levellevel, group, partition, sex, analyte level, material level, concentration level
Day, Run, Replicateday; run; replicate
Unit, Analyte, Matrixunit; analyte; matrix
Instrumentinstrument, device, analyzer, system
Excludedexcluded
Exclusion reasonexclusion reason, exclusion_reason, reason
Expected or assigned concentrationexpected, expected concentration, expected result, nominal concentration, assigned concentration, concentration
Lotlot, reagent lot, reagent_lot, calibrator lot, calibrator_lot
Specimen volumespecimen volume, volume specimen, volume_specimen, sample volume
Diluent volumediluent volume, volume diluent, volume_diluent
Stated dilution factordilution factor, dilution_factor, factor, dilution, stated factor
Serial stepsserial steps, serial_steps, steps
Diluentdiluent
Sequence positionposition, sequence position, order, sequence
Rolerole, sample role, aliquot role, material role
Interferent concentrationinterferent concentration, interferent conc, interferent_conc, interferent
Elapsed timetimepoint, time point, time, elapsed time, storage time
Storage conditioncondition, storage condition, handling condition, storage

Descriptive summary, repeatability and bias against an assigned value#

Chapters: Descriptive summary, Repeatability, Bias against an assigned value.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSpecimen / material IDExcept result-only descriptive importsCoded ID of the specimen or control material; descriptive imports generate IDs when this column is absent
ResultResultYesThe measured value, in the study unit
Level or groupLevelNoConcentration or control level. Each level is summarized separately
DayDayNoDay label
RunRunNoRun label
ReplicateReplicateNoReplicate label

Repeatability expects one material, day and run for each level. For bias, enter each level's assigned value on Set up, and record Day and Run: repeats of one material get a mean and bias interval only when they share one recorded day and run.

Multi-day precision#

Chapter: Multi-day precision.

Same fields as above. Day, Run and Replicate are also required, though the dialog doesn't mark them. Use one material for each level and at least two days. Every day needs the same number of runs, and every run the same number of replicates, at least two.

Reference interval verification and establishment#

Chapters: Reference interval verification, Reference interval establishment.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSubject IDYesCoded ID of the reference subject. A subject with repeat draws is flagged and counted once
ResultResultYesThe measured value
Level or groupPartitionNoThe partition, for example female or male. Each partition is analyzed separately. Without it, all rows form partition 1
ReplicateReplicateNoReplicate label

Day and Run can be imported but aren't shown in the grid.

Linearity#

Chapter: Linearity.

Column useGrid columnRequiredWhat to enter
Expected (assigned) valueExpected concentrationYesThe prepared concentration, zero or more
ResultResultYesThe measured value
Level or groupLevelNoLevel label
Specimen or material IDSpecimen IDNoID of the preparation
Day, Runnot shownNoLabels
ReplicateReplicateNoReplicate label

Dilution verification#

Chapter: Dilution verification.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSpecimen IDYesCoded ID of the high specimen
ResultMeasured resultYesThe measured post-dilution result
Specimen volumeSpecimen volumeSee ruleSpecimen volume, in the same unit as the diluent volume
Diluent volumeDiluent volumeSee ruleVolume of diluent
Dilution factorStated factorSee ruleOverall dilution factor, for example 5. Use 1 for undiluted material
Serial dilution stepsSerial stepsSee ruleStepwise factors, for example 5, 2, or 5; 2 in a decimal-comma study. The overall factor is their product
DiluentDiluentNoName of the diluent
Level or groupLevelNoLevel label
ReplicateReplicateNoReplicate label

Each row needs both volumes, a stated factor or serial steps. If you give more than one, they must agree. Give a factor as a number, for example 5 for a 1-in-5 dilution.

Method comparison#

Chapter: Method comparison.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSpecimen IDYesCoded specimen ID
Comparative method resultComparative method (x)YesResult by the comparative method
Test method resultTest method (y)YesResult by the test method, on the same specimen
Level or groupSpecimen group (level)NoA subgroup label
ReplicateReplicateNoReplicate label. Replicate pairs of one specimen are averaged

Put both results for a specimen in the same row.

Instrument comparison#

Chapter: Instrument comparison.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSpecimen IDYesCoded specimen ID, repeated on each instrument's row
InstrumentInstrumentYesThe instrument that produced this result
ResultResultYesThe measured value
Level or groupLevelNoLevel label
ReplicateReplicateNoReplicate label

Enter one row for each result. The instrument list on Set up is filled from the file. Unit, Analyte and Matrix still group the file.

Reagent lot comparison#

Chapter: Lot comparison.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSpecimen IDYesCoded specimen ID. An ID beginning with QC (any case) marks quality-control material
ResultResultYesThe measured value
LotLotYesMust match the Previous lot or New lot on Set up exactly
Level or groupLevelNoLevel label
Day, Run, ReplicateDay, Run, ReplicateNoLabels

Qualitative agreement#

Chapter: Qualitative agreement.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSpecimen IDYesCoded specimen ID
Comparative method resultComparative method resultYesA declared category or nonreportable label
Test method resultTest method resultYesA declared category or nonreportable label
ReplicateReplicateNoReplicate label. The first included row for a specimen is its result

Set the allowed values on Set up. By default the result categories are negative and positive, and the nonreportable labels are equivocal, invalid and indeterminate. Case is ignored. See Qualitative results that are not reportable.

Near-cutoff precision#

Chapter: Near-cutoff precision.

Column useGrid columnRequiredWhat to enter
Specimen or material IDSample IDYesCoded sample ID
Expected (assigned) valueConcentrationYesThe sample concentration, above zero
ResultResult categoryYesA declared category or nonreportable label
DayDayNoDay label
ReplicateReplicateNoReplicate label

Detection limits#

Chapter: Detection limits.

Column useGrid columnRequiredWhat to enter
Specimen or material IDMaterial IDYesCoded ID of the blank or low-level material. Give each material one role and one assigned concentration
Sample roleRoleYesblank (analyte-free material) or low (low-level material), any case
ResultResultYesThe measured value. Blanks often give negative values. Enter <x only with the rank-below setting
Expected (assigned) valueAssigned concentrationFor LoQ low materialsRequired and above zero for low materials in a LoQ study
LotReagent lotNoLot label
Day, Run, ReplicateDay, Run, ReplicateNoLabels

Carryover#

Chapter: Carryover.

Column useGrid columnRequiredWhat to enter
RunRunYesRun or sequence ID. Each run gets its own decision
Position in the runSequence positionYesA whole number, such as 3, used once in each run. It sets the order
Sample roleRoleYeshigh, low, H or L, any case
ResultResultYesThe measured value
Specimen or material IDMaterial IDNoID of the low or high material

In the 21-position design, each run has positions 1 to 21 in the order L L L H H L H H L L L L H H L H H L H H L, from one low and one high material.

Interference#

Chapter: Interference.

Column useGrid columnRequiredWhat to enter
Specimen or material IDPool IDYesShared by the base, control and test aliquots of one preparation
Sample roleRoleYesbase (neat specimen), control (vehicle-matched, no interferent) or test (with interferent), any case
Interferent concentrationInterferent concentrationYesFor test aliquots, the interferent concentration, above 0, in the unit set on Set up. For base and control aliquots, leave it empty or enter 0
ResultResultYesThe measured analyte value
Level or groupAnalyte levelNoLabel such as low or high. Each level is analyzed separately
ReplicateReplicateNoReplicate label

Stability#

Chapter: Stability.

Column useGrid columnRequiredWhat to enter
Specimen or material IDMaterial / specimen IDYesCoded ID
Time pointElapsed timeYesA number in the Time unit set on Set up, for example 24 for 24 h. Baseline rows use the Baseline time, 0 by default
ResultResultYesThe measured value
Storage conditionStorage conditionFor non-baseline rowsA condition set on Set up, any case. Baseline rows can leave it blank
ReplicateReplicateNoReplicate label

Sigma metric#

Chapter: Sigma metric.

Enter the concentration, allowable total error, bias and imprecision in the Data stage, or copy them from other studies' results.