Each chapter starts with Run this study, a one-page guide to take to the bench. To start a study, choose New study, enter the analyte and unit, pick the study type and choose Create study. The tables below give each study type's name and group in that list.

Before collecting results#

Set your acceptance limits before you collect. A specimen measured five times is one specimen with five replicates; the guides count both.

Collect what the guide's suggested plan gives. The calculation minimum is the least data the software can calculate from, not a collection plan. The example downloads show the file layout only.

Methods and sources explains the calculations, their sources and how they were checked.

The four steps every study follows#

Set up → Data → Results → Report. Each stage tab shows whether it needs input, is ready, is done or is out of date. The main button in the study header names the next step.

  1. Set up. Read the Collection plan, check Study details, set Acceptance limits and fill in the study type's own settings. See Study setup.
  2. Data. Choose Import a file or Add row. Blank template (CSV) and Run this study are at the top of the stage. In a row's Correct / exclude column, open the row to correct a value or to exclude it (an exclusion needs a reason). See Entering data.
  3. Results. Choose Calculate results, or Recalculate results after a change. Results opens with the verdict and Why. See Calculating results.
  4. Report. Read the report, then choose Download PDF or Print. A report prints with blank Signed off by and Date lines to sign by hand; Sign off fills them in. See The study report.

Precision, bias and linearity#

ChapterStudy type in New studyTypical data
Descriptive summaryDescriptive summary under SummariesAny list of results
RepeatabilityRepeatability under PrecisionOne material measured repeatedly in one run
Multi-day precisionMulti-day precision under PrecisionMaterial, day, run, replicate, result
Bias against an assigned valueBias check under Comparison and biasRepeated results on a material with an assigned value
LinearityLinearity under Measuring rangeExpected concentration and measured result
Dilution verificationDilution verification under Measuring rangeMeasured result with volumes or dilution factor
Derived sigmaSigma metric under SummariesTEa, bias and imprecision at each concentration

Method, instrument and lot comparison#

ChapterStudy type in New studyTypical data
Method comparisonMethod comparison under Comparison and biasSpecimen, comparative result (x), test result (y)
Instrument comparisonInstrument comparison under Comparison and biasSpecimen, instrument, result
Reagent lot comparisonReagent lot comparison under Comparison and biasSpecimen, lot, result
Qualitative agreementQualitative agreement under Qualitative testsSpecimen, comparative result, test result

Reference ranges and detection limits#

ChapterStudy type in New studyTypical data
Reference interval verificationVerify a reference interval under Reference intervalsSubject, result
Reference interval establishmentEstablish a reference interval under Reference intervalsSubject, result, optional partition
Detection limitsDetection limits under Measuring rangeMaterial, role (blank or low), result
Near-cutoff precisionNear-cutoff precision under Qualitative testsSample, concentration, result category

Stability, carryover and interference#

ChapterStudy type in New studyTypical data
StabilityStability under Sample and material effectsMaterial, storage condition, elapsed time, result
CarryoverCarryover under Sample and material effectsRun, sequence position, role, material, result
InterferenceInterference under Sample and material effectsPool, role, interferent concentration, result

Before you choose#

For an ADM QC export, use QC targets (ADM export) in the Tools menu to prepare targets and compare instruments from your QC records.

  • A study's design is fixed once it has data or a calculation. For a different question, start a new study. See Choosing a study.