Each study answers one question. The New study dialog lists the studies by topic and sets up the matching study type. If you already have a results file, Import a file can suggest the study instead; see importing a file.

Starting a study#

  1. Open a project and select New study. In an empty project, select Or choose a study first. In an empty workspace, the Projects page opens on Start a study instead: select a study, then Start this study, and the dialog opens with that study chosen.
  2. Fill in Study title (optional), Analyte and Unit. Opened from a project, the dialog fills in the analyte and unit most of its studies use.
  3. Under Study type, select a study, or type part of its name in Find a study type…. Its question, Suggested collection and what You get appear below the list, with a Run this study link to its guide. Some studies ask a few more details there (below); these can also wait until Set up.
  4. Select:
    • Create study to open the study at Set up, where you check the collection plan and record your acceptance limits before collecting, or
    • Create and import data if you already have results. A sigma study has only Create study. In an empty workspace, or when you pick New project in the dialog, a project named after the analyte, for example Glucose evaluation, is made for the study.

Which study answers your question#

Your questionChooseChapter
How much do results vary within runs, between runs and between days?Multi-day precisionMulti-day precision
How closely do repeated results of one material agree within a single run?RepeatabilityRepeatability
Do two methods or instruments give the same results on the same patient samples?Method comparisonMethod comparison
Do two or more instruments agree when your file has one result per row?Instrument comparisonInstrument comparison
Does a new reagent lot give the same patient results as the current lot?Reagent lot comparisonLot comparison
How far are results from the assigned value of a reference or control material?Bias checkBias against an assigned value
Is the response a straight line across the measuring range?LinearityLinearity
Does a diluted high sample recover its expected value after correction?Dilution verificationDilution verification
What is the lowest amount the method can detect or quantify?Detection limitsDetection limits
Does hemolysis or another added substance change results?InterferenceInterference
Does a high sample raise the result of the low sample run after it?CarryoverCarryover
How long do specimens, reagents, calibrators or controls stay stable in storage?StabilityStability
Do healthy-subject results fit an existing reference interval or limit?Verify a reference intervalReference interval verification
What reference interval or limit do your healthy-subject results support?Establish a reference intervalReference interval establishment
Do positive and negative results agree with a comparative method?Qualitative agreementQualitative agreement
How reliably does a qualitative test call samples near its cutoff?Near-cutoff precisionNear-cutoff precision
What sigma do your known bias and imprecision give against the allowable total error?Sigma metricSigma metric
What are the mean, SD and CV of a set of results?Descriptive summaryDescriptive summary

Once a study exists, both reference-interval choices show the study type Reference interval, and both qualitative choices show Qualitative agreement or near-cutoff precision.

What each form asks#

Every study asks for Analyte and Unit and an optional Study title. Qualitative agreement needs only Analyte. A blank title becomes, for example, Glucose · repeatability. Some studies ask more:

StudyExtra questions
Method comparisonComparative method or instrument, Test method or instrument
Instrument comparisonReference analyzer
Bias checkAssigned value
Dilution verificationLowest reportable value before dilution, Highest reportable value before dilution, Expected value of the undiluted sample
Reagent lot comparisonPrevious lot name, New lot name, Required matched pairs. Set each specimen's allowable error and the concentration groups on Set up.
Verify a reference intervalReference limits (two-sided or one-sided), Claimed lower reference limit, Claimed upper reference limit
Establish a reference intervalReference limits (two-sided or one-sided)
Detection limitsWhich limit (limit of blank, detection or quantitation)
InterferenceInterferent, Interferent concentration unit
StabilityTime unit, What is stored, Same material measured at every time?, Largest acceptable change from baseline (%) (optional)

Type numbers in these fields with a decimal point (1.5). The study's Blank template (CSV) on Set up and Data has the columns it needs; your own columns can have any names and order.

After you create the study#

Create study opens the study at Set up, with its collection plan open. Create and import data opens the study at Data with Import a file open. A carryover study also offers Start a blank 21-position run on Data, and a sigma study takes values instead of a data table.

If the calculation needs a setting the dialog didn't ask for, such as a lot study's allowable error or the quantitation goal for LoQ, the study header shows Finish set up and Set up names what's missing. You can change everything later on Set up.

The study design is fixed once data exist#

You can change Study type only until the study has rows or a saved calculation. After that, create a new study and import the same file again. To remove the old study, choose More, then Move study to Trash; see deleting a study.

Start again with the same setup#

  1. Open the study you want to reuse. Choose More → New study from this setup. Any unsaved changes are saved first.
  2. Enter a Study title, choose the Project and check the Study date. Select Create study.
  3. Check the new study's Set up stage before collecting or importing results. The study details, calculation settings, assigned values and acceptance limits are copied. Enter the new lot names and check any settings that depend on this experiment.

The new study starts with empty data and its own version history. Its source study, saved calculations and reports are unchanged. Reagent and calibrator lots, the previous and new lots of a lot comparison, and links to earlier calculations are cleared; select the precision evidence again if the new comparison needs a linked Deming lambda. A sigma study starts with an empty entries table.